“A practical method to determine the effective resolution in incoherent experimental electron tomography”. Heidari Mezerji H, van den Broek W, Bals S, Ultramicroscopy 111, 330 (2011). http://doi.org/10.1016/j.ultramic.2011.01.021
Abstract: It is not straightforward to determine resolution for a 3D reconstruction when performing an electron tomography experiment. Different contributions such as missing wedge and misalignment add up and often influence the final resolution in an anisotropic manner. The conventional resolution measures can not be used for all of the reconstruction techniques, especially for iterative techniques which are more commonly used for electron tomography in materials science. Here we define a quantitative resolution measure that determines the resolution in three orthogonal directions of the reconstruction. As an application we use this measure to determine the optimum number of simultaneous iterative reconstruction technique (SIRT) iterations to reconstruct the gold nanoparticles, based on a high angle annular dark field STEM (HAADF-STEM) tilt series.
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT); Vision lab
Impact Factor: 2.843
Times cited: 26
DOI: 10.1016/j.ultramic.2011.01.021
|
|
|
“Precision of three-dimensional atomic scale measurements from HRTEM images : what are the limits?”.Wang A, Van Aert S, Goos P, van Dyck D, Ultramicroscopy 114, 20 (2012). http://doi.org/10.1016/j.ultramic.2011.12.002
Abstract: In this paper, we investigate to what extent high resolution transmission electron microscopy images can be used to measure the mass, in terms of thickness, and surface profile, corresponding to the defocus offset, of an object at the atomic scale. Therefore, we derive an expression for the statistical precision with which these object parameters can be estimated in a quantitative analysis. Evaluating this expression as a function of the microscope settings allows us to derive the optimal microscope design. Acquiring three-dimensional structure information in terms of thickness turns out to be much more difficult than obtaining two-dimensional information on the projected atom column positions. The attainable precision is found to be more strongly affected by processes influencing the image contrast, such as phonon scattering, than by the specific choice of microscope settings. For a realistic incident electron dose, it is expected that atom columns can be distinguished with single atom sensitivity up to a thickness of the order of the extinction distance. A comparable thickness limit is determined to measure surface steps of one atom. An increase of the electron dose shifts the limiting thickness upward due to an increase in the signal-to-noise ratio.
Keywords: A1 Journal article; Engineering Management (ENM); Electron microscopy for materials research (EMAT); Vision lab
Impact Factor: 2.843
Times cited: 5
DOI: 10.1016/j.ultramic.2011.12.002
|
|
|
“The properties of SIRT, TVM, and DART for 3D imaging of tubular domains in nanocomposite thin-films and sections”. Chen D, Goris B, Bleichrodt F, Heidari Mezerji H, Bals S, Batenburg KJ, de With G, Friedrich H, Ultramicroscopy 147, 137 (2014). http://doi.org/10.1016/j.ultramic.2014.08.005
Abstract: In electron tomography, the fidelity of the 3D reconstruction strongly depends on the employed reconstruction algorithm. In this paper, the properties of SIRT, TVM and DART reconstructions are studied with respect to having only a limited number of electrons available for imaging and applying different angular sampling schemes. A well-defined realistic model is generated, which consists of tubular domains within a matrix having slab-geometry. Subsequently, the electron tomography workflow is simulated from calculated tilt-series over experimental effects to reconstruction. In comparison with the model, the fidelity of each reconstruction method is evaluated qualitatively and quantitatively based on global and local edge profiles and resolvable distance between particles. Results show that the performance of all reconstruction methods declines with the total electron dose. Overall, SIRT algorithm is the most stable method and insensitive to changes in angular sampling. TVM algorithm yields significantly sharper edges in the reconstruction, but the edge positions are strongly influenced by the tilt scheme and the tubular objects become thinned. The DART algorithm markedly suppresses the elongation artifacts along the beam direction and moreover segments the reconstruction which can be considered a significant advantage for quantification. Finally, no advantage of TVM and DART to deal better with fewer projections was observed.
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT)
Impact Factor: 2.843
Times cited: 42
DOI: 10.1016/j.ultramic.2014.08.005
|
|
|
“Quantification of crystalline and amorphous content in porous TiO2 samples from electron energy loss spectroscopy”. Bertoni G, Beyers E, Verbeeck J, Mertens M, Cool P, Vansant EF, Van Tendeloo G, Ultramicroscopy 106, 630 (2006). http://doi.org/10.1016/j.ultramic.2006.03.006
Abstract: We present an efficient method for the quantification of crystalline versus amorphous phase content in mesoporous materials, making use of electron energy loss spectroscopy. The method is based on fitting a superposition of core-loss edges using the maximum likelihood method with measured reference spectra. We apply the method to mesoporous TiO2 samples. We show that the absolute amount of the crystalline phase can be determined with an accuracy below 5%. This method takes also the amorphous phase into account, where standard X-ray diffraction is only quantitative for crystalline phases and not for amorphous phase. (c) 2006 Elsevier B.V.. All rights reserved.
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT); Laboratory of adsorption and catalysis (LADCA)
Impact Factor: 2.843
Times cited: 83
DOI: 10.1016/j.ultramic.2006.03.006
|
|
|
“Quantitative atomic resolution mapping using high-angle annular dark field scanning transmission electron microscopy”. Van Aert S, Verbeeck J, Erni R, Bals S, Luysberg M, van Dyck D, Van Tendeloo G, Ultramicroscopy 109, 1236 (2009). http://doi.org/10.1016/j.ultramic.2009.05.010
Abstract: A model-based method is proposed to relatively quantify the chemical composition of atomic columns using high angle annular dark field (HAADF) scanning transmission electron microscopy (STEM) images. The method is based on a quantification of the total intensity of the scattered electrons for the individual atomic columns using statistical parameter estimation theory. In order to apply this theory, a model is required describing the image contrast of the HAADF STEM images. Therefore, a simple, effective incoherent model has been assumed which takes the probe intensity profile into account. The scattered intensities can then be estimated by fitting this model to an experimental HAADF STEM image. These estimates are used as a performance measure to distinguish between different atomic column types and to identify the nature of unknown columns with good accuracy and precision using statistical hypothesis testing. The reliability of the method is supported by means of simulated HAADF STEM images as well as a combination of experimental images and electron energy-loss spectra. It is experimentally shown that statistically meaningful information on the composition of individual columns can be obtained even if the difference in averaged atomic number Z is only 3. Using this method, quantitative mapping at atomic resolution using HAADF STEM images only has become possible without the need of simultaneously recorded electron energy loss spectra.
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT); Vision lab
Impact Factor: 2.843
Times cited: 166
DOI: 10.1016/j.ultramic.2009.05.010
|
|
|
“Real space maps of atomic transitions”. Schattschneider P, Verbeeck J, Hamon AL, Ultramicroscopy 109, 781 (2009). http://doi.org/10.1016/j.ultramic.2009.01.016
Abstract: Considering the rapid technical development of transmission electron microscopes, we investigate the possibility to map electronic transitions in real space on the atomic scale. To this purpose, we analyse the information carried by the scatterer's initial and final state wave functions and the role of the different atomic transition channels for the inelastic scattering cross section. It is shown that the change in the magnetic quantum number in the transition can be mapped. Two experimental set-ups are proposed, one blocking half the diffraction plane, the other one using a cylinder lens for imaging. Both methods break the conventional circular symmetry in the electron microscope making it possible to detect the handedness of electronic transitions as an asymmetry in the image intensity. This finding is of important for atomic resolution energy-loss magnetic chiral dichroism (EMCD), allowing to obtain the magnetic moments of single atoms.
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT)
Impact Factor: 2.843
Times cited: 10
DOI: 10.1016/j.ultramic.2009.01.016
|
|
|
“Real space maps of magnetic moments on the atomic scale: theory and feasibility”. Schattschneider P, Ennen I, Stoger-Pollach M, Verbeeck J, Mauchamp V, Jaouen M, Ultramicroscopy 110, 1038 (2010). http://doi.org/10.1016/j.ultramic.2009.11.020
Abstract: The recently discovered EMCD technique (energy loss magnetic chiral dichroism) can detect atom specific magnetic moments with nanometer resolution, exploiting the spin selectivity of electronic transitions in energy loss spectroscopy. Yet, direct imaging of magnetic moments on the atomic scale is not possible. In this paper we present an extension of EMCD that can overcome this limit. As a model system we chose bcc Fe. We present image simulations of the L3 white line signal, based on the kinetic equation for the density matrix of the 200 kV probe electron. With actual progress in instrumentation (high brightness sources, aberration corrected lenses) this technique should allow direct imaging of spin moments on the atomic scale.
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT)
Impact Factor: 2.843
Times cited: 10
DOI: 10.1016/j.ultramic.2009.11.020
|
|
|
“The reciprocal space of carbon tubes: a detailed interpretation of the electron diffraction effects”. Zhang XB, Zhang XF, Amelinckx S, Van Tendeloo G, van Landuyt J, Ultramicroscopy 54, 237 (1994). http://doi.org/10.1016/0304-3991(94)90123-6
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT)
Impact Factor: 2.436
Times cited: 59
DOI: 10.1016/0304-3991(94)90123-6
|
|
|
“Refinement of the 200 structure factor for GaAs using parallel and convergent beam electron nanodiffraction data”. Müller K, Schowalter M, Jansen J, Tsuda K, Titantah J, Lamoen D, Rosenauer A, Ultramicroscopy 109, 802 (2009). http://doi.org/10.1016/j.ultramic.2009.03.029
Abstract: We present a new method to measure structure factors from electron spot diffraction patterns recorded under almost parallel illumination in transmission electron microscopes. Bloch wave refinement routines have been developed to refine the crystal thickness, its orientation and structure factors by comparison of experimentally recorded and calculated intensities. Our method requires a modicum of computational effort, making it suitable for contemporary personal computers. Frozen lattice and Bloch wave simulations of GaAs diffraction patterns are used to derive optimised experimental conditions. Systematic errors are estimated from the application of the method to simulated diffraction patterns and rules for the recognition of physically reasonable initial refinement conditions are derived. The method is applied to the measurement of the 200 structure factor for GaAs. We found that the influence of inelastically scattered electrons is negligible. Additionally, we measured the 200 structure factor from zero loss filtered two-dimensional convergent beam electron diffraction patterns. The precision of both methods is found to be comparable and the results agree well with each other. A deviation of more than 20% from isolated atom scattering data is observed, whereas close agreement is found with structure factors obtained from density functional theory [A. Rosenauer, M. Schowalter, F. Glas, D. Lamoen, Phys. Rev. B 72 (2005), 085326-1], which account for the redistribution of electrons due to chemical bonding via modified atomic scattering amplitudes.
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT)
Impact Factor: 2.843
Times cited: 8
DOI: 10.1016/j.ultramic.2009.03.029
|
|
|
“Selective imaging of sublattices in complex structures”. Amelinckx S, Milat O, Van Tendeloo G, Ultramicroscopy 51, 90 (1993). http://doi.org/10.1016/0304-3991(93)90138-N
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT)
Impact Factor: 2.436
Times cited: 8
DOI: 10.1016/0304-3991(93)90138-N
|
|
|
“Selective imaging of the “substructures&rdquo, in incommensurately modulated intergrowth crystal structures”. Milat O, Van Tendeloo G, Amelinckx S, Ultramicroscopy 41, 65 (1992). http://doi.org/10.1016/0304-3991(92)90095-2
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT)
Impact Factor: 2.436
Times cited: 5
DOI: 10.1016/0304-3991(92)90095-2
|
|
|
“A simple preparation method for air-sensitive specimens for transmission electron microscopy demonstrated by Rb6C60”. Zhang XF, Zhang XB, Bernaerts D, Van Tendeloo G, Amelinckx S, van Landuyt J, Werner H, Ultramicroscopy 55, 25 (1994). http://doi.org/10.1016/0304-3991(94)90077-9
Abstract: In this paper a particularly simple but efficient method is presented by which samples of alkali-doped C-60 materials or other air-sensitive materials can be prepared and transferred into a transmission electron microscope for direct observations and investigations. Flexible, transparent glove bags are used which are filled to a slight overpressure with dry nitrogen. Under this protective atmosphere, the air-sensitive sample is mounted in the specimen holder and inserted in the vacuum of the electron microscope. Rb6C60 which is prepared and transferred into the microscope in this way has been investigated by transmission electron microscopy (TEM). The results confirm the bcc structure and especially the location of the rubidium atoms.
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT)
Impact Factor: 2.436
Times cited: 2
DOI: 10.1016/0304-3991(94)90077-9
|
|
|
“Strain measurement at the nanoscale : comparison between convergent beam electron diffraction, nano-beam electron diffraction, high resolution imaging and dark field electron holography”. Béché, A, Rouviere JL, Barnes JP, Cooper D, Ultramicroscopy 131, 10 (2013). http://doi.org/10.1016/j.ultramic.2013.03.014
Abstract: Convergent beam electron diffraction (CBED), nano-beam electron diffraction (NBED or NBD), high resolution imaging (HRTEM and HRSTEM) and dark field electron holography (DFEH or HoloDark) are five TEM based techniques able to quantitatively measure strain at the nanometer scale. In order to demonstrate the advantages and disadvantages of each technique, two samples composed of epitaxial silicon-germanium layers embedded in a silicon matrix have been investigated. The five techniques are then compared in terms of strain precision and accuracy, spatial resolution, field of view, mapping abilities and ease of performance and analysis. (C) 2013 Elsevier By. All rights reserved.
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT)
Impact Factor: 2.843
Times cited: 73
DOI: 10.1016/j.ultramic.2013.03.014
|
|
|
“Structures and phase transitions in C60 and C70 fullerites”. Van Tendeloo G, Amelinckx S, Muto S, Verheijen MA, van Loosdrecht PHM, Meijer G, Ultramicroscopy 51, 168 (1993). http://doi.org/10.1016/0304-3991(93)90145-N
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT)
Impact Factor: 2.436
Times cited: 17
DOI: 10.1016/0304-3991(93)90145-N
|
|
|
“TEM sample preparation by FIB for carbon nanotube interconnects”. Ke X, Bals S, Romo Negreira A, Hantschel T, Bender H, Van Tendeloo G, Ultramicroscopy 109, 1353 (2009). http://doi.org/10.1016/j.ultramic.2009.06.011
Abstract: A powerful method to study carbon nanotubes (CNTs) grown in patterned substrates for potential interconnects applications is transmission electron microscopy (TEM). However, high-quality TEM samples are necessary for such a study. Here, TEM specimen preparation by focused ion beam (FIB) has been used to obtain lamellae of patterned samples containing CNTs grown inside contact holes. A dual-cap Pt protection layer and an extensive 5 kV cleaning procedure are applied in order to preserve the CNTs and avoid deterioration during milling. TEM results show that the inner shell structure of the carbon nanotubes has been preserved, which proves that focused ion beam is a useful technique to prepare TEM samples of CNT interconnects.
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT)
Impact Factor: 2.843
Times cited: 21
DOI: 10.1016/j.ultramic.2009.06.011
|
|
|
“The evolution of HVEM application in antwerp”. van Landuyt J, Ultramicroscopy
T2 –, 2nd Osaka International Symp.on High-Voltage Electron Microscopy : New Directions and Future Aspects of High Voltage Electron Microscopy, November 8-10, 1990, Osaka University, Osaka, Japan 39, 287 (1991). http://doi.org/10.1016/0304-3991(91)90208-N
Abstract: The evolution of the use of the 1250 keV high-voltage electron microscope in Antwerp is sketched by illustrating a non-exhaustive set of examples in various fields. One of the main present fields of application gets some more attention, i.e. the defect studies as produced by processing steps in microelectronic devices: (i) strain-induced dislocations at the edges of various device isolation interlayers, (ii) morphologies resulting from high-energy ion implantation creating buried layers for silicon on insulator (SOI) and other implantation technologies.
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT)
Impact Factor: 2.436
DOI: 10.1016/0304-3991(91)90208-N
|
|
|
“The Fresnel effect of a defocused biprism on the fringes in inelastic holography”. Verbeeck J, Bertoni G, Schattschneider P, Ultramicroscopy
T2 –, 16th International Microscopy Congress, SEP 03-08, 2006, Sapporo, JAPAN 108, 263 (2008). http://doi.org/10.1016/j.ultramic.2007.06.007
Abstract: We present energy filtered holography experiments on a thin foil of Al. By propagating the reduced density matrix of the probe electron through the microscope, we quantitatively predict the fringe contrast as a function of energy loss. Fringe contrast simulations include the effect of Fresnel fringes created at the edges of the defocused biprism, the effect of partial coherence in combination with inelastic scattering, and the effect of a finite energy distribution of the incoming beam. (c) 2007 Elsevier B.V. All rights reserved.
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT)
Impact Factor: 2.843
Times cited: 15
DOI: 10.1016/j.ultramic.2007.06.007
|
|
|
“Theory of free electron vortices”. Schattschneider P, Verbeeck J, Ultramicroscopy 111, 1461 (2011). http://doi.org/10.1016/j.ultramic.2011.07.004
Abstract: The recent creation of electron vortex beams and their first practical application motivates a better understanding of their properties. Here, we develop the theory of free electron vortices with quantized angular momentum, based on solutions of the Schrödinger equation for cylindrical boundary conditions. The principle of transformation of a plane wave into vortices with quantized angular momentum, their paraxial propagation through round magnetic lenses, and the effect of partial coherence are discussed.
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT)
Impact Factor: 2.843
Times cited: 57
DOI: 10.1016/j.ultramic.2011.07.004
|
|
|
“Throughput maximization of particle radius measurements by balancing size and current of the electron probe”. van den Broek W, Van Aert S, Goos P, van Dyck D, Ultramicroscopy 111, 940 (2011). http://doi.org/10.1016/j.ultramic.2010.11.025
Abstract: In thispaperweinvestigatewhichprobesizemaximizesthethroughputwhenmeasuringtheradiusof nanoparticlesinhighangleannulardarkfieldscanningtransmissionelectronmicroscopy(HAADFSTEM). The sizeandthecorrespondingcurrentoftheelectronprobedeterminetheprecisionoftheestimateofa particlesradius.Maximizingthroughputmeansthatamaximumnumberofparticlesshouldbeimaged withinagiventimeframe,sothataprespecifiedprecisionisattained.WeshowthatBayesianstatistical experimentaldesignisaveryusefulapproachtodeterminetheoptimalprobesizeusingacertainamount of priorknowledgeaboutthesample.Thedependenceoftheoptimalprobesizeonthedetectorgeometry and thediameter,variabilityandatomicnumberoftheparticlesisinvestigated.Anexpressionforthe optimalprobesizeintheabsenceofanykindofpriorknowledgeaboutthespecimenisderivedaswell.
Keywords: A1 Journal article; Engineering Management (ENM); Electron microscopy for materials research (EMAT); Vision lab
Impact Factor: 2.843
Times cited: 7
DOI: 10.1016/j.ultramic.2010.11.025
|
|
|
“Tomographic spectroscopic imaging, an experimental proof of concept”. van den Broek W, Verbeeck J, Schryvers D, de Backer S, Scheunders P, Ultramicroscopy 109, 296 (2009). http://doi.org/10.1016/j.ultramic.2008.11.022
Abstract: Recording the electron energy loss spectroscopy data cube with a series of energy filtered images is a dose inefficient process because the energy slit blocks most of the electrons. When recording the data cube by scanning an electron probe over the sample, perfect dose efficiency is attained; but due to the low current in nanoprobes, this often is slower, with a smaller field of view. In W. Van den Broek et al. [Ultramicroscopy, 106 (2006) 269], we proposed a new method to record the data cube, which is more dose efficient than an energy filtered series. It produces a set of projections of the data cube and then tomographically reconstructs it. In this article, we demonstrate these projections in practice, we present a simple geometrical model that allows for quantification of the projection angles and we present the first successful experimental reconstruction, all on a standard post-column instrument.
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT); Vision lab
Impact Factor: 2.843
Times cited: 1
DOI: 10.1016/j.ultramic.2008.11.022
|
|
|
“Dark field electron holography for strain measurement”. Béché, A, Rouvière JL, Barnes JP, Cooper D, Ultramicroscopy 111, 227 (2011). http://doi.org/10.1016/J.ULTRAMIC.2010.11.030
Abstract: Dark field electron holography is a new TEM-based technique for measuring strain with nanometer scale resolution. Here we present the procedure to align a transmission electron microscope and obtain dark field holograms as well as the theoretical background necessary to reconstruct strain maps from holograms. A series of experimental parameters such as biprism voltage, sample thickness, exposure time, tilt angle and choice of diffracted beam are then investigated on a silicon-germanium layer epitaxially embedded in a silicon matrix in order to obtain optimal dark field holograms over a large field of view with good spatial resolution and strain sensitivity.
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT)
Impact Factor: 2.843
Times cited: 31
DOI: 10.1016/J.ULTRAMIC.2010.11.030
|
|
|
“Bandgap measurement of high refractive index materials by off-axis EELS”. Vatanparast M, Egoavil R, Reenaas TW, Verbeeck J, Holmestad R, Vullum PE, Ultramicroscopy 182, 92 (2017). http://doi.org/10.1016/J.ULTRAMIC.2017.06.019
Abstract: In the present work Cs aberration corrected and monochromated scanning transmission electron microscopy electron energy loss spectroscopy (STEM-EELS) has been used to explore experimental setups that allow bandgaps of high refractive index materials to be determined. Semi-convergence and collection angles in the mu rad range were combined with off-axis or dark field EELS to avoid relativistic losses and guided light modes in the low loss range to contribute to the acquired EEL spectra. Off-axis EELS further supressed the zero loss peak and the tail of the zero loss peak. The bandgap of several GaAs-based materials were successfully determined by simple regression analyses of the background subtracted EEL spectra. The presented set-up does not require that the acceleration voltage is set to below the. Cerenkov limit and can be applied over the entire acceleration voltage range of modern TEMs and for a wide range of specimen thicknesses. (C) 2017 Elsevier B.V. All rights reserved.
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT)
Impact Factor: 2.843
Times cited: 3
DOI: 10.1016/J.ULTRAMIC.2017.06.019
|
|
|
“Optimization of NBED simulations for disc-detection measurements”. Grieb T, Krause FF, Mahr C, Zillmann D, Müller-Caspary K, Schowalter M, Rosenauer A, Ultramicroscopy 181, 50 (2017). http://doi.org/10.1016/J.ULTRAMIC.2017.04.015
Abstract: Nano-beam electron diffraction (NBED) is a method which can be applied to measure lattice strain and polarisation fields in strained layer heterostructures and transistors. To investigate precision, accuracy and spatial resolution of such measurements in dependence of properties of the specimen as well as electron optical parameters, simulations of NBED patterns are required which allow to predict the result of common disc-detection algorithms. In this paper we demonstrate by focusing on the detection of the central disc in crystalline silicon that such simulations require to take several experimental characteristics into account in order to obtain results which are comparable to those from experimental NBED patterns. These experimental characteristics are the background intensity, the presence of Poisson noise caused by electron statistics and blurring caused by inelastic scattering and by the transfer quality of the microscope camera. By means of these optimized simulations, different effects of specimen properties on disc detection – such as strain, surface morphology and compositional changes on the nanometer scale – are investigated and discussed in the context of misinterpretation in experimental NBED evaluations. It is shown that changes in surface morphology and chemical composition lead to measured shifts of the central disc in the NBED pattern of tens to hundreds of grad. These shifts are of the same order of magnitude or even larger than shifts that could be caused by an electric polarisation field in the range of MV/cm. (C) 2017 Elsevier B.V. All rights reserved.
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT)
Impact Factor: 2.843
Times cited: 6
DOI: 10.1016/J.ULTRAMIC.2017.04.015
|
|
|
“Frozen lattice and absorptive model for high angle annular dark field scanning transmission electron microscopy : a comparison study in terms of integrated intensity and atomic column position measurement”. Alania M, Lobato Hoyos IP, Van Aert S, Ultramicroscopy 184, 188 (2018). http://doi.org/10.1016/J.ULTRAMIC.2017.08.021
Abstract: <script type='text/javascript'>document.write(unpmarked('In this paper, both the frozen lattice (FL) and the absorptive potential (AP) approximation models are compared in terms of the integrated intensity and the precision with which atomic columns can be located from an image acquired using high angle annular dark field (HAADF) scanning transmission electron microscopy (STEM). The comparison is made for atoms of Cu, Ag, and Au. The integrated intensity is computed for both an isolated atomic column and an atomic column inside an FCC structure. The precision has been computed using the so-called Cramer-Rao Lower Bound (CRLB), which provides a theoretical lower bound on the variance with which parameters can be estimated. It is shown that the AP model results into accurate measurements for the integrated intensity only for small detector ranges under relatively low angles and for small thicknesses. In terms of the attainable precision, both methods show similar results indicating picometer range precision under realistic experimental conditions. (C) 2017 Elsevier B.V. All rights reserved.'));
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT)
Impact Factor: 2.843
DOI: 10.1016/J.ULTRAMIC.2017.08.021
|
|
|
“Quantitative HAADF STEM of SiGe in presence of amorphous surface layers from FIB preparation”. Grieb T, Tewes M, Schowalter M, Müller-Caspary K, Krause FF, Mehrtens T, Hartmann J-M, Rosenauer A, Ultramicroscopy 184, 29 (2018). http://doi.org/10.1016/J.ULTRAMIC.2017.09.012
Abstract: <script type='text/javascript'>document.write(unpmarked('The chemical composition of four Si1-xGex layers grown on silicon was determined from quantitative scanning transmission electron microscopy (STEM). The chemical analysis was performed by a comparison of the high-angle annular dark field (HAADF) intensity with multislice simulations. It could be shown that amorphous surface layers originating from the preparation process by focused-ion beam (FIB) at 30 kV have a strong influence on the quantification: the local specimen thickness is overestimated by approximately a factor of two, and the germanium concentration is substantially underestimated. By means of simulations, the effect of amorphous surface layers on the HAADF intensity of crystalline silicon and germanium is investigated. Based on these simulations, a method is developed to analyze the experimental HAADF-STEM images by taking the influence of the amorphous layers into account which is done by a reduction of the intensities by multiplication with a constant factor. This suggested modified HAADF analysis gives germanium concentrations which are in agreement with the nominal values. The same TEM lamella was treated with low-voltage ion milling which removed the amorphous surface layers completely. The results from subsequent quantitative HAADF analyses are in agreement with the nominal concentrations which validates the applicability of the used frozen-lattice based multislice simulations to describe the HAADF scattering of Si1-xGex in STEM. (C) 2017 Elsevier B.V. All rights reserved.'));
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT)
Impact Factor: 2.843
Times cited: 7
DOI: 10.1016/J.ULTRAMIC.2017.09.012
|
|
|
“Exploring possibilities of band gap measurement with off-axis EELS in TEM”. Korneychuk S, Partoens B, Guzzinati G, Ramaneti R, Derluyn J, Haenen K, Verbeeck J, Ultramicroscopy 189, 76 (2018). http://doi.org/10.1016/J.ULTRAMIC.2018.03.021
Abstract: A technique to measure the band gap of dielectric materials with high refractive index by means of energy electron loss spectroscopy (EELS) is presented. The technique relies on the use of a circular (Bessel) aperture and suppresses Cherenkov losses and surface-guided light modes by enforcing a momentum transfer selection. The technique also strongly suppresses the elastic zero loss peak, making the acquisition, interpretation and signal to noise ratio of low loss spectra considerably better, especially for excitations in the first few eV of the EELS spectrum. Simulations of the low loss inelastic electron scattering probabilities demonstrate the beneficial influence of the Bessel aperture in this setup even for high accelerating voltages. The importance of selecting the optimal experimental convergence and collection angles is highlighted. The effect of the created off-axis acquisition conditions on the selection of the transitions from valence to conduction bands is discussed in detail on a simplified isotropic two band model. This opens the opportunity for deliberately selecting certain transitions by carefully tuning the microscope parameters. The suggested approach is experimentally demonstrated and provides good signal to noise ratio and interpretable band gap signals on reference samples of diamond, GaN and AlN while offering spatial resolution in the nm range. (C) 2018 Elsevier B.V. All rights reserved.
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT); Condensed Matter Theory (CMT)
Impact Factor: 2.843
Times cited: 7
DOI: 10.1016/J.ULTRAMIC.2018.03.021
|
|
|
“Strain analysis from nano-beam electron diffraction : influence of specimen tilt and beam convergence”. Grieb T, Krause FF, Schowalter M, Zillmann D, Sellin R, Müller-Caspary K, Mahr C, Mehrtens T, Bimberg D, Rosenauer A, Ultramicroscopy 190, 45 (2018). http://doi.org/10.1016/J.ULTRAMIC.2018.03.013
Abstract: Strain analyses from experimental series of nano-beam electron diffraction (NBED) patterns in scanning transmission electron microscopy are performed for different specimen tilts. Simulations of NBED series are presented for which strain analysis gives results that are in accordance with experiment. This consequently allows to study the relation between measured strain and actual underlying strain. A two-tilt method which can be seen as lowest-order electron beam precession is suggested and experimentally implemented. Strain determination from NBED series with increasing beam convergence is performed in combination with the experimental realization of a probe-forming aperture with a cross inside. It is shown that using standard evaluation techniques, the influence of beam convergence on spatial resolution is lower than the influence of sharp rings around the diffraction disc which occur at interfaces and which are caused by the tails of the intensity distribution of the electron probe. (C) 2018 Elsevier B.V. All rights reserved.
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT)
Impact Factor: 2.843
Times cited: 1
DOI: 10.1016/J.ULTRAMIC.2018.03.013
|
|
|
“Comparison of first moment STEM with conventional differential phase contrast and the dependence on electron dose”. Müller-Caspary K, Krause FF, Winkler F, Béché, A, Verbeeck J, Van Aert S, Rosenauer A, Ultramicroscopy 203, 95 (2019). http://doi.org/10.1016/J.ULTRAMIC.2018.12.018
Abstract: This study addresses the comparison of scanning transmission electron microscopy (STEM) measurements of momentum transfers using the first moment approach and the established method that uses segmented annular detectors. Using an ultrafast pixelated detector to acquire four-dimensional, momentum-resolved STEM signals, both the first moment calculation and the calculation of the differential phase contrast (DPC) signals are done for the same experimental data. In particular, we investigate the ability to correct the segment-based signal to yield a suitable approximation of the first moment for cases beyond the weak phase object approximation. It is found that the measurement of momentum transfers using segmented detectors can approach the first moment measurement as close as 0.13 h/nm in terms of a root mean square (rms) difference in 10 nm thick SrTiO3 for a detector with 16 segments. This amounts to 35% of the rms of the momentum transfers. In addition, we present a statistical analysis of the precision of first moment STEM as a function of dose. For typical experimental settings with recent hardware such as a Medipix3 Merlin camera attached to a probe-corrected STEM, we find that the precision of the measurement of momentum transfers stagnates above certain doses. This means that other instabilities such as specimen drift or scan noise have to be taken into account seriously for measurements that target, e.g., the detection of bonding effects in the charge density.
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT)
Impact Factor: 2.843
Times cited: 25
DOI: 10.1016/J.ULTRAMIC.2018.12.018
|
|
|
“Efficient first principles simulation of electron scattering factors for transmission electron microscopy”. Susi T, Madsen J, Ludacka U, Mortensen JJ, Pennycook TJ, Lee Z, Kotakoski J, Kaiser U, Meyer JC, Ultramicroscopy 197, 16 (2019). http://doi.org/10.1016/J.ULTRAMIC.2018.11.002
Abstract: Electron microscopy is a powerful tool for studying the properties of materials down to their atomic structure. In many cases, the quantitative interpretation of images requires simulations based on atomistic structure models. These typically use the independent atom approximation that neglects bonding effects, which may, however, be measurable and of physical interest. Since all electrons and the nuclear cores contribute to the scattering potential, simulations that go beyond this approximation have relied on computationally highly demanding all-electron calculations. Here, we describe a new method to generate ab initio electrostatic potentials when describing the core electrons by projector functions. Combined with an interface to quantitative image simulations, this implementation enables an easy and fast means to model electron scattering. We compare simulated transmission electron microscopy images and diffraction patterns to experimental data, showing an accuracy equivalent to earlier all-electron calculations at a much lower computational cost.
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT)
Impact Factor: 2.843
Times cited: 3
DOI: 10.1016/J.ULTRAMIC.2018.11.002
|
|
|
“High dose efficiency atomic resolution imaging via electron ptychography”. Pennycook TJ, Martinez GT, Nellist PD, Meyer JC, Ultramicroscopy 196, 131 (2019). http://doi.org/10.1016/J.ULTRAMIC.2018.10.005
Abstract: Radiation damage places a fundamental limitation on the ability of microscopy to resolve many types of materials at high resolution. Here we evaluate the dose efficiency of phase contrast imaging with electron ptychography. The method is found to be far more resilient to temporal incoherence than conventional and spherical aberration optimized phase contrast imaging, resulting in significantly greater clarity at a given dose. This robustness is explained by the presence of achromatic lines in the four dimensional ptychographic dataset.
Keywords: A1 Journal article; Electron microscopy for materials research (EMAT)
Impact Factor: 2.843
Times cited: 1
DOI: 10.1016/J.ULTRAMIC.2018.10.005
|
|