Abstract |
The contamination of ground samples by a commercially available Lovibond McCrone Micronizing Mill is discussed. Tracer and weighing experiments showed that abrasion of corundum grinding elements was important, introducing 620 mg of abrasion products per minute of wet grinding. Agate grinding elements were abraded at ⩽6 mg min-1. The abrasion products and grinding elements were analyzed by x-ray fluorescence, spark-source mass spectrometry and neutron activation analysis. Contamination in trace element analysis of geological materials is likely to be negligible for agate grinding elements and, except for a few transition metals, also for corundum grinding elements. Contamination of typical biological samples is significant for a few elements even when agate elements are used, and is absolutely prohibitive for trace analysis when corundum elements are used. |
|