Abstract |
Selenium is determined in the ng g−1 to μg g−1 range in biological and environmental samples. A wet digestion procedure was optimized with respect to volatility losses and recovery yields, by using 75Se metabolically incorporated into rat organs. Selenium is preconcentrated from the digestion liquid by a two-step reduction with 4 M HCl and ascorbic acid. The colloidal selenium formed is adsorbed on activated carbon and filtered on a Nucleopore membrane for measurement by energy-dispersive x-ray fluorescence. Almost complete recovery was obtained, and the detection limit was 20 ng, corresponding to 10 ng g−1 for a 2-g sample. Biological reference materials were analyzed with satisfactory results, and the accuracy of the method was good. |
|