Abstract: A comparison of four methods for the determination of barium, lanthanum, cerium and neodymium in lateritic materials from Brazil is presented. Three of the methods were based on x-ray fluorescence (XRF) spectroscopy: two by radioisotope excitation (Co-57 and Am-241) and one by secondary target XRF (a molybdenum target with a tungsten anode). The other method was based on neutron activation analysis employing both a Ge(Li) coaxial detector and a high-purity germanium detector. The results from these four methods were similar for lanthanum, cerium and neodymium, but for barium at low concentrations (<500 ppm) the neutron activation and the secondary target XRF methods were not suitable. Data on the precision and accuracy of these methods using a series of standard reference rocks are given. The advantages and limitations of each of these methods with respect to the analysis of lateritic materials are discussed.
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
DOI: 10.1002/XRS.1300150105