Abstract: The methodology of direct hair analysis by energy-dispersive x-ray fluorescence was studied. The effect on the XRF result of having a non-homogeneous radial distribution of the analyte in a single hair strand and the macroscopic effects in a bundle of hairs were calculated to evaluate possible systematic errors. The detection limits were mapped as a function of the target thickness and surface fraction. It appeared that a 10 mg cm−2 sample thickness, i. e. a target with about four layers of hair strands, is recommended. The standard deviation of this simple direct analytical method is 619% for some important elements. Discrepancies with neutron activation analysis had a mean value of around 15%. About twelve elements can be determined simultaneously on a routine basis.
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
DOI: 10.1002/XRS.1300130106