“A novel gas inlet system for improved aerosol entrainment in laser ablation inductively coupled plasma mass spectrometry”. Bleiner D, Altorfer H, Journal of analytical atomic spectrometry (2005). http://doi.org/10.1039/B505248C
Abstract: In order to minimize the dead volume in large cells for laser ablation inductively coupled plasma mass spectrometry, and improve the aerosol entrainment characteristics, the gas inlet nozzle has been set in rotation. This allowed a wider volume to be swept than with the traditional static inlet nozzle approach. Therefore, sensitivity combined with site-to-site repeatability was improved by a factor of two, together with minimization of aerosol loss within the cell and signal dispersion.
Keywords: A1 Journal article; Plasma Lab for Applications in Sustainability and Medicine – Antwerp (PLASMANT)
Impact Factor: 3.379
Times cited: 21
DOI: 10.1039/B505248C
|
“Overcoming pulse mixing and signal tailing in laser ablation inductively coupled plasma mass spectrometry depth profiling”. Bleiner D, Belloni F, Doria D, Lorusso A, Nassisi V, Journal of analytical atomic spectrometry (2005). http://doi.org/10.1039/B509379C
Abstract: The laser ablation-induced plasma was used as a composition-con trolled source for ion implantation in Si crystals. Then, laser ablation in combination with inductively coupled plasma mass spectrometry was used for the elemental depth profiling of the implanted samples. Monte Carlo simulations permitted us to conclude that a depth resolution of tens of nm would be necessary to define the shape of the implantation profiles, as is obtained using XPS and RBS, whereas a hundred nm depth resolution is sufficient to determine the total implanted dose. The detection power of LA-ICP-MS would routinely allow rapid analytical control on the trace level implanted dose. Nevertheless, this technique is limited in terms of depth profiling resolution due to pulse mixing and signal tailing induced during the aerosol transport. Raw signal processing procedures were developed for the minimization of shapeline dispersion, deconvolution of pulse mixing and more appropriate assessment of the implanted profiles. Shapeline dispersion could be corrected for by determining the signal waning constant and implementing this information for a non-affine alibi transformation of the LA-ICP-MS signal traces. Pulse mixing deconvolution was attained with an algorithm that considered accumulated signal intensity due to pulse-on-pulse stacking, i.e., the latest pulse on top of all antecedent individual pulses' exponential tails proportionally.
Keywords: A1 Journal article; Plasma Lab for Applications in Sustainability and Medicine – Antwerp (PLASMANT)
Impact Factor: 3.379
Times cited: 26
DOI: 10.1039/B509379C
|
“Plasma diagnostics and numerical simulations: insight into the heart of analytical glow discharges”. Bogaerts A, Journal of analytical atomic spectrometry 22, 13 (2007). http://doi.org/10.1039/b611436a
Keywords: A1 Journal article; Plasma Lab for Applications in Sustainability and Medicine – Antwerp (PLASMANT)
Impact Factor: 3.379
Times cited: 23
DOI: 10.1039/b611436a
|
“Revealing hidden paint layers in oil paintings by means of scanning macro-XRF : a mock-up study based on Rembrandt's “An old man in military costume””. Alfeld M, de Nolf W, Cagno S, Janssens K, et al, Journal of analytical atomic spectrometry 28, 40 (2013). http://doi.org/10.1039/C2JA30119A
Abstract: Over the past several decades the oeuvre of Rembrandt has been the subject of extensive art historical and scientific investigations. One of the most striking features to emerge is his frequent re-use of canvases and panels. The painting An Old Man in Military Costume (78.PB.246), in the collection of the J. Paul Getty Museum, is an example of such a re-used panel. Conventional imaging techniques revealed the presence of a second portrait under the surface portrait, but the details of this hidden portrait have not yet been revealed. Vermilion (HgS) has been identified to have been used nearly exclusively in the flesh tones of the lower painting, suggesting that element-specific XRF imaging might successfully image the hidden portrait. To test this hypothesis, a full-scale mock-up of the painting was created, including a “free impression” of the hidden portrait, reproducing as closely as possible the pigments and paint stratigraphy of the original painting. XRF imaging of the mock-up painting was conducted using three different XRF imaging systems: a mobile X-ray tube based system and two synchrotron-based setups (one equipped with multiple SDDs and one equipped with a Maia detector). The sensitivity, limits of detection and imaging capabilities of each system under the chosen experimental conditions are evaluated and compared. The results indicate that an investigation of the original painting by this method would have an excellent chance of success.
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
Impact Factor: 3.379
DOI: 10.1039/C2JA30119A
|
“Quantitative analysis of zirconium oxide by direct glow discharge mass spectrometry using a secondary cathode”. Schelles W, Van Grieken R, Journal of analytical atomic spectrometry 12, 49 (1997)
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
|
“Analysis of speleothems by electron and X-ray microprobes”. Kuczumow A, Vekemans B, Schalm O, Gysels K, Ro C-U, Van Grieken R, Journal of analytical atomic spectrometry 16, 90 (2001). http://doi.org/10.1039/B007725I
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
DOI: 10.1039/B007725I
|
“Characteristics of atomic absorption calibration curves with the transversely heated graphite furnace”. Bencs L, Szakács O, Szoboszlai N, Ajtony Z, Bozsai G, Journal of analytical atomic spectrometry 18, 105 (2003). http://doi.org/10.1039/B208198A
Abstract: Calibration curves with two quasi-linear sections ("double sloping'') were observed for the medium volatile elements, Cr and Cu, with the use of a SIMAA 6000 graphite furnace atomic absorption spectrometer under interrupted internal gas flow conditions. If a standard transversally heated graphite atomizer (THGA) tube was shortened by 0.5 mm at both of its ends, (i.e. the gaps were enlarged between graphite furnace housing and tube ends), a stronger declination of the calibration curves resulted. Elements with fairly high diffusion coefficients (>5.8 cm(2) s(-1)) and with short appearance time of their transients (<0.6 s), such as Cr and Cu, have shown the most characteristic sensitivity drop towards higher concentrations. This anomalous feature could be eliminated in three different ways; (1) by applying end-capped THGA tubes, (2) using mini-flow (50 cm(3) min(-1)) conditions during the atomization stage, and (3) by adding Pd-Mg chemical modifier. For the low volatile Mo and V, the calibration curves had no irregular shape. For Ag, Co, Cr, Cu, Mn and Ni, the mini-flow settings improved the linearity of the calibration curves and extended the upper limit of the linear calibration range by a factor of 1.5-2.0. The irregular characteristic of the analytical curves was interpreted as an increased vapour loss at higher analyte concentrations through the opened ends of the standard THGA tubes. This vapour loss was associated with the significantly diverse expulsion velocities of atoms, caused by the difference in temperature and concentration gradients, when evaporating amounts of analytes with more than one order of magnitude difference.
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
DOI: 10.1039/B208198A
|
“Development of a multi-method analytical approach based on the combination of synchrotron radiation X-ray micro-analytical techniques and vibrational micro-spectroscopy methods to unveil the causes and mechanism of darkening of “fake-gilded&rdquo, decorations in a Cimabue painting”. Monico L, Prati S, Sciutto G, Catelli E, Romani A, Balbas DQ, Li Z, De Meyer S, Nuyts G, Janssens K, Cotte M, Garrevoet J, Falkenberg G, Tardillo Suarez VI, Tucoulou R, Mazzeo R, Journal of analytical atomic spectrometry 37, 114 (2022). http://doi.org/10.1039/D1JA00271F
Abstract: Redox processes activated by environmental factors have been identified as the main cause of the chromatic alterations of a number of artists' pigments, including the yellow pigment orpiment (As2S3). Although a general comprehension of the mechanisms has been provided through characterization of degradation compounds of As2S3, experimental evidences to prove how other paint components and how different environmental agents influence the formation pathways of specific secondary compounds are still lacking. Thus, it becomes fundamental to develop a methodological strategy which enable achieving a discrimination among the causes affecting the chemical stability of more heterogenous As2S3-based paints and defining the mechanism through which the alteration establishes and evolves, with the ultimate goal of optimizing the preventive conservation measures of unique masterpieces. In this paper, we propose a comprehensive multi-material and multi-method approach based on the combination of synchrotron radiation X-ray micro-analytical techniques (i.e., X-ray diffraction, X-ray fluorescence and X-ray absorption near edge structure spectroscopy at S K-/Ag L-3-/As K-edges) and vibrational micro-spectroscopy methods to unveil the causes and mechanism of darkening of “fake-gilded” decorations in tempera paintings, originally consisting of an unusual mixture of As2S3 and metallic silver (Ag-0). Such degradation process is a not yet understood phenomenon threatening a series of Old Master paintings, including those by the Italian painters Cimabue and Pietro Lorenzetti. The high specificity, sensitivity and lateral resolution of the employed analytical methods allowed providing first-time evidence for the presence of black acanthite (alpha-Ag2S), mimetite [Pb-5(AsO4)(3)Cl] and syngenite [K2Ca(SO4)(2)center dot H2O] as degradation products of the “fake-gilded” decorations in the Maesta by Cimabue (Church of Santa Maria dei Servi, Bologna, Italy). Furthermore, the study of the painting combined with that of tempera paint mock-ups permitted to explore and define the environmental agents and internal factors causing the darkening, by proving that: (i) Ag-0 and moisture are key-factors for triggering the transformation of As2S3 to alpha-Ag2S and As-oxides; (ii) S2--ions arising from the degradation of As2S3 are the main responsible for the formation of alpha-Ag2S; (iii) light exposure strengthens the tendency of the paint components towards alteration. Based on our findings, we finally propose a degradation mechanism of As2S3/Ag-0-based tempera paints.
Keywords: A1 Journal article; Antwerp X-ray Imaging and Spectroscopy (AXIS)
Impact Factor: 3.4
DOI: 10.1039/D1JA00271F
|
“Microscopic X-ray fluorescence analysis”. Janssens K, Vincze L, Rubio J, Bernasconi G, Adams F, Journal of analytical atomic spectrometry 9, 151 (1994). http://doi.org/10.1039/JA9940900151
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
DOI: 10.1039/JA9940900151
|
“Suitability of polycapillary optics for focusing of monochromatic synchrotron radiation as used in trace level micro-XANES measurements”. Vincze L, Wei F, Proost K, Vekemans B, Janssens K, He Y, Yan Y, Falkenberg G, Journal of analytical atomic spectrometry 17, 177 (2002). http://doi.org/10.1039/B110210A
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
Impact Factor: 3.379
Times cited: 23
DOI: 10.1039/B110210A
|
“Inductively coupled plasma-mass spectrometry: insights through computer modeling”. Bogaerts A, Aghaei M, Journal of analytical atomic spectrometry 32, 233 (2017). http://doi.org/10.1039/C6JA00408C
Abstract: In this tutorial review paper, we illustrate how computer modeling can contribute to a better insight in inductively coupled plasma-mass spectrometry (ICP-MS). We start with a brief overview on previous efforts, studying the fundamentals of the ICP and ICP-MS, with main focus on previous modeling activities. Subsequently, we explain in detail the model that we developed in previous years, and we show typical calculation results, illustrating the plasma characteristics, gas flow patterns and the sample transport, evaporation and ionization. We also present the effect of various experimental parameters, such as operating conditions, geometrical aspects and sample characteristics, to illustrate how modeling can help to elucidate the optimal conditions for improved analytical performance.
Keywords: A1 Journal article; Plasma Lab for Applications in Sustainability and Medicine – Antwerp (PLASMANT)
Impact Factor: 3.379
Times cited: 14
DOI: 10.1039/C6JA00408C
|
“Inorganic nitrogen speciation in single micrometer-size particles by laser microprobe mass analysis”. Bruynseels F, Otten P, Van Grieken R, Journal of analytical atomic spectrometry 3, 237 (1988). http://doi.org/10.1039/JA9880300237
Abstract: Individual micrometer-size particles of inorganic nitrogen salts, such as NaNO3, NaNO2, (NH4)2SO4, NH4Cl, NH4NO3 and their mixtures, were analysed in the laser microprobe mass analyser. The cluster-ion distributions of the salts are systematically described and can be used to gain speciation information. The results can be used to study the distribution of the atmospheric aerosol inorganic nitrogen compounds and to investigate their chemical transformations under ambient conditions.
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
DOI: 10.1039/JA9880300237
|
“Modeling of a microsecond pulsed glow discharge: behavior of the argon excited levels and of the sputtered copper atoms and ions”. Bogaerts A, Gijbels R, Journal of analytical atomic spectrometry 16, 239 (2001). http://doi.org/10.1039/b009289o
Keywords: A1 Journal article; Plasma Lab for Applications in Sustainability and Medicine – Antwerp (PLASMANT)
Impact Factor: 3.379
Times cited: 36
DOI: 10.1039/b009289o
|
“Occurrence of gas flow rotational motion inside the ICP torch : a computational and experimental study”. Aghaei M, Flamigni L, Lindner H, Günther D, Bogaerts A, Journal of analytical atomic spectrometry 29, 249 (2014). http://doi.org/10.1039/c3ja50302j
Abstract: An inductively coupled plasma, connected to the sampling cone of a mass spectrometer, is computationally investigated. The occurrence of rotational motion of the auxiliary and carrier gas flows is studied. The effects of operating parameters, i.e., applied power and gas flow rates, as well as geometrical parameters, i.e., sampler orifice diameter and injector inlet diameter, are investigated. Our calculations predict that at higher applied power the auxiliary and carrier gas flows inside the torch move more forward to the sampling cone, which is validated experimentally for the auxiliary gas flow, by means of an Elan 6000 ICP-MS. Furthermore, an increase of the gas flow rates can also modify the occurrence of rotational motion. This is especially true for the carrier gas flow rate, which has a more pronounced effect to reduce the backward motion than the flow rates of the auxiliary and cooling gas. Moreover, a larger sampler orifice (e.g., 2 mm instead of 1 mm) reduces the backward flow of the auxiliary gas path lines. Finally, according to our model, an injector inlet of 2 mm diameter causes more rotations in the carrier gas flow than an injector inlet diameter of 1.5 mm, which can be avoided again by changing the operating parameters.
Keywords: A1 Journal article; Plasma Lab for Applications in Sustainability and Medicine – Antwerp (PLASMANT)
Impact Factor: 3.379
Times cited: 21
DOI: 10.1039/c3ja50302j
|
“Micro and surface analysis in archaeology”. Adams F, Adriaens A, Aerts A, de Raedt I, Janssens K, Schalm O, Journal of analytical atomic spectrometry 12, 257 (1997). http://doi.org/10.1039/A606091I
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
Impact Factor: 3.379
DOI: 10.1039/A606091I
|
“Diffusion- and velocity-driven spatial separation of analytes from single droplets entering an ICP off-axis”. Borovinskaya O, Aghaei M, Flamigni L, Hattendorf B, Tanner M, Bogaerts A, Günther D, Journal of analytical atomic spectrometry 29, 262 (2014). http://doi.org/10.1039/c3ja50307k
Abstract: The reproducible temporal separation of ion signals generated from a single multi-element droplet, observed in previous studies, was investigated in detail in this work using an ICPTOFMS with high temporal resolution. It was shown that the signal peak intensities of individual elements temporally shift relative to each other only for droplets moving through the plasma off-axis. The magnitude of these shifts correlated with the vaporization temperatures of the analytes and depended on the radial position of the droplets as well as on the thermal properties and velocity profiles of the carrier gases of the ICP. The occurrence of the signal shifting was explained by a spatial separation of analytes already present in the vapor phase in the ICP from a yet unvaporized residue of the droplet. This separation is most likely driven by anisotropic diffusion of vaporized analytes towards the plasma axis and a radial velocity gradient. The proposed explanation is supported by modeling of the gas velocities inside the ICP and imaging of the atomic and ionic emissions produced from single droplets, whose patterns were sloping towards the center of the torch. The effects observed in these studies are important not only for the fundamental understanding of analyteplasma interactions but have also a direct impact on the signal intensities and stability.
Keywords: A1 Journal article; Plasma Lab for Applications in Sustainability and Medicine – Antwerp (PLASMANT)
Impact Factor: 3.379
Times cited: 18
DOI: 10.1039/c3ja50307k
|
“Numerical simulation analysis of flow patterns and particle transport in the HEAD laser ablation cell with respect to inductively coupled plasma spectrometry”. Lindner H, Autrique D, Pisonero J, Günther D, Bogaerts A, Journal of analytical atomic spectrometry 25, 295 (2010). http://doi.org/10.1039/b920905k
Abstract: The present study analyses a specific laser ablation cell, the High Efficiency Aerosol Dispersion (HEAD) cell (see J. Pisonero et al., J. Anal. At. Spectrom., 2006, 21, 922931), by means of computational fluid dynamics (CFD) simulations. However, this cell consists of different modular parts, therefore, the results are probably of interest for the further development of other ablation cells. In the HEAD cell, the ablation spot is positioned below an orifice in the ceiling of the sample chamber. The particle transport through this orifice has been analysed for a ceiling height of 0.8 mm. The critical velocity for the onset of particle losses was found to be independent on the ejection angle at the crater spot. The deceleration of the particles can be described as the stopping in an effectively steady gas. Particle losses were negligible in this modular part of the cell at the evaluated laser ablation conditions. The transport efficiency through the Venturi chamber was investigated for different sample gas flow rates. In this case, it was found that small particles were predominantly lost at low flow rates, the large particles at higher flow rates. Making use of the simulation results, it was possible to design a modification of the HEAD cell that results in extremely short calculated washout times. The simulations yielded a signal of less than 10 ms, which was produced by more than 99% of the introduced mass.
Keywords: A1 Journal article; Plasma Lab for Applications in Sustainability and Medicine – Antwerp (PLASMANT)
Impact Factor: 3.379
Times cited: 16
DOI: 10.1039/b920905k
|
“Microscopical X-ray fluorescence analysis and related methods with laboratory and synchrotron radiation sources”. Adams F, Janssens K, Snigirev A, Journal of analytical atomic spectrometry 13, 319 (1998). http://doi.org/10.1039/A707100K
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
Impact Factor: 3.379
DOI: 10.1039/A707100K
|
“The colour of silver stained glass : analytical investigations carried out with XRF, SEM/EDX, TEM and IBA”. Jembrih-Simbürger D, Neelmeijer C, Schalm O, Fredrickx P, Schreiner M, De Vis K, Mäder M, Schryvers D, Caen J, Journal of analytical atomic spectrometry 17, 321 (2002). http://doi.org/10.1039/b111024c
Abstract: Glass treated on its surface with silver compounds and an aluminosilicate, such as ochre or clay, at higher temperatures (between 550 and 650 °C) accepts a wide variety of a yellow colour. It is the aim of this study to investigate the parameters of the manufacturing process affecting the final colour of silver stained glass and to correlate them with the final colour and colour intensity. Therefore, defined mixtures of ochre and a silver compound (AgCl, AgNO3, Ag2SO4, Ag3PO4, Ag2O) were prepared and applied on soda-lime glass. The firing process was modified within the range from 563 to 630 °C and glass samples were analysed after treatment with energy dispersive X-ray fluorescence analysis (EDXRF), scanning electron microscopy (SEM/EDX), transmission electron microscopy (TEM), as well as ion beam analysis (IBA) with an external beam. Within the scope of IBA simultaneous measurements using particle-induced X-ray emission (PIXE), particle-induced gamma-ray emission (PIGE), and Rutherford backscattering spectrometry (RBS) were carried out in order to obtain the thickness of the Ag-rich surface layer and the depth distribution of Ag. By means of TEM the microstructure of the silver particles was visualised. XRF results show that the lowest amount of Ag could be detected on glass samples treated with silver stain mixtures containing AgCl and Ag2O. A low kiln temperature (e.g. 563 °C) results in a higher silver concentration at the surface and lower penetration depths. Furthermore, the results obtained with SEM/EDX at cross-sections of the glass samples could be confirmed by PIXE, PIGE, RBS, and TEM.
Keywords: A1 Journal article; Art; Electron microscopy for materials research (EMAT); AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
Impact Factor: 3.379
Times cited: 42
DOI: 10.1039/b111024c
|
“The use of lead-glass capillaries for microfocusing of highly energetic (0-60 KeV) synchrotron radiation”. Janssens K, Vincze L, Vekemans B, Adams F, Haller M, Knöchel A, Journal of analytical atomic spectrometry 13, 339 (1998). http://doi.org/10.1039/A707700I
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
Impact Factor: 3.379
DOI: 10.1039/A707700I
|
“A large area full-field EDXRF imaging system based on a THCOBRA gaseous detector”. Silva ALM, Carvalho ML, Janssens K, Veloso JFCA, Journal of analytical atomic spectrometry 30, 343 (2015). http://doi.org/10.1039/C4JA00301B
Abstract: By taking advantage of the attractive features in terms of gain, position and energy resolution of the 2D-THCOBRA detector, a new large area Full-Field X-ray Fluorescence Imaging (FF-XRFI) system for EDXRF imaging applications was developed. The proposed FF-XRFI system has an active area of 10 x 10 cm(2) and can be used to examine macroscopic samples with a moderately good energy resolution (< 1.6 keV FWHM at 8 keV) and a suitable spatial resolution (similar to 500 mu m FWHM). This combination of characteristics allows us to record elemental distribution maps from the surface of different sample types by combining image and energy information. Two different approaches were used for X-ray optics, one based on a single pinhole and another based on a parallel multiple-hole collimator. To illustrate the system capabilities, some sample examples were imaged and studied.
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
Impact Factor: 3.379
Times cited: 11
DOI: 10.1039/C4JA00301B
|
“Temporal and spatially resolved laser-scattering plasma diagnostics for the characterization of a ms-pulsed glow discharge”. Gamez G, Bogaerts A, Hieftje GM, Journal of analytical atomic spectrometry 21, 350 (2006). http://doi.org/10.1039/b511764j
Keywords: A1 Journal article; Plasma Lab for Applications in Sustainability and Medicine – Antwerp (PLASMANT)
Impact Factor: 3.379
Times cited: 17
DOI: 10.1039/b511764j
|
“Micro-heterogeneity study of trace elements in USGS, MPI-DING and glass reference materials by means of synchrotron micro-XRF”. Kempenaers L, Janssens K, Jochum KP, Vincze L, Vekemans B, Somogyi A, Drakopoulos M, Adams F, Journal of analytical atomic spectrometry 18, 350 (2003). http://doi.org/10.1039/B212196D
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
Impact Factor: 3.379
Times cited: 38
DOI: 10.1039/B212196D
|
“Laser ablation of copper in different background gases: comparative study by numerical modeling and experiments”. Bogaerts A, Chen Z, Bleiner D, Journal of analytical atomic spectrometry 21, 384 (2006). http://doi.org/10.1039/b514313f
Keywords: A1 Journal article; Plasma Lab for Applications in Sustainability and Medicine – Antwerp (PLASMANT)
Impact Factor: 3.379
Times cited: 67
DOI: 10.1039/b514313f
|
“Quantification problems in light element determination by grazing emission X-ray fluorescence”. Kuczumow A, Claes M, Schmeling M, Van Grieken R, de Gendt S, Journal of analytical atomic spectrometry 15, 415 (2000). http://doi.org/10.1039/A908724I
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
DOI: 10.1039/A908724I
|
“Surface analysis of halide distributions in complex AgX microcrystals by imaging time-of-flight SIMS (TOF-SIMS)”. Verlinden G, Gijbels R, Geuens I, de Keyzer R, Journal of analytical atomic spectrometry 14, 429 (1999). http://doi.org/10.1039/a807276k
Keywords: A1 Journal article; Plasma Lab for Applications in Sustainability and Medicine – Antwerp (PLASMANT)
Impact Factor: 3.379
Times cited: 10
DOI: 10.1039/a807276k
|
“Effects of adding hydrogen to an argon glow discharge: overview of relevant processes and some qualitative explanations”. Bogaerts A, Gijbels R, Journal of analytical atomic spectrometry 15, 441 (2000). http://doi.org/10.1039/a909779a
Keywords: A1 Journal article; Plasma Lab for Applications in Sustainability and Medicine – Antwerp (PLASMANT)
Impact Factor: 3.379
Times cited: 58
DOI: 10.1039/a909779a
|
“The ID21 X-ray and infrared microscopy beamline at the ESRF: status and recent applications to artistic materials”. Cotte M, Pouyet E, Salome M, Rivard C, De Nolf W, Castillo-Michel H, Fabris T, Monico L, Janssens K, Wang T, Sciau P, Verger L, Cormier L, Dargaud O, Brun E, Bugnazet D, Fayard B, Hesse B, del Real AEP, Veronesi G, Langlois J, Balcar N, Vandenberghe Y, Sole VA, Kieffer J, Barrett R, Cohen C, Cornu C, Baker R, Gagliardini E, Papillon E, Susini J, Journal of analytical atomic spectrometry 32, 477 (2017). http://doi.org/10.1039/C6JA00356G
Abstract: The ID21 beamline (European Synchrotron Radiation facility, France) is a multi micro-analytical platform combining X-ray and infrared micro-probes, for characterization of elements, species, molecular groups and crystalline structures in complex materials. Applications are mainly in the fields of cultural heritage, life science, environmental and earth sciences, materials sciences. Here, we first present the status of instruments: (i) the scanning micro-spectroscopy end-station, operating from 2.0 to 9.2 keV, under vacuum and offering cryo conditions, for the acquisition of 2D micro X-ray fluorescence (mu XRF) maps, single point micro X-ray Absorption Near Edge Structure (mu XANES) spectra and speciation maps with sub-micrometric resolution; (ii) the XANES full-field end-station, operating in the same vacuum and energy conditions, for the acquisition of hyper-spectral radiographs of thin concentrated samples, resulting in speciation maps with micrometric resolution and millimetric field of view; (iii) the scanning micro-X-ray diffraction (mu XRD)/mu XRF end-station, operating at 8.5 keV, in air, for the acquisition of 2D crystalline phase maps, with micrometric resolution; and (iv) the scanning infrared microscope, operating in the mid-infrared range for the acquisition of molecular maps and some structural maps with micrometric resolution. Recent hardware and software developments are presented, as well as new protocols for improved sample preparation of thin sections. Secondly, a review of recent applications for the study of cultural heritage is presented, illustrated by various examples: determination of the origin of the color in blue Chinese porcelains and in brown Sevres porcelains; detection of lead in ink on Herculaneum papyri; identification and degradation of modeling materials used by Auguste Rodin and of chrome yellow pigments used by Vincent van Gogh. Cryo capabilities are illustrated by the analysis of plants exposed to chromate solutions. These examples show the variety of materials analyzed, of questions tackled, and particularly the multiple advantages of the ID21 analytical platform for the analysis of ancient and artistic materials.
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
Impact Factor: 3.379
Times cited: 39
DOI: 10.1039/C6JA00356G
|
“Characterization of AgxAuy nano particles by TEM and STEM”. de Vyt A, Gijbels R, Davock H, van Roost C, Geuens I, Journal of analytical atomic spectrometry 14, 499 (1999). http://doi.org/10.1039/a807695b
Keywords: A1 Journal article; Plasma Lab for Applications in Sustainability and Medicine – Antwerp (PLASMANT)
Impact Factor: 3.379
Times cited: 2
DOI: 10.1039/a807695b
|
“The afterglow mystery of pulsed glow discharges and the role of dissociative electron-ion recombination”. Bogaerts A, Journal of analytical atomic spectrometry 22, 502 (2007). http://doi.org/10.1039/b618035c
Keywords: A1 Journal article; Plasma Lab for Applications in Sustainability and Medicine – Antwerp (PLASMANT)
Impact Factor: 3.379
Times cited: 56
DOI: 10.1039/b618035c
|