“Diffusion- and velocity-driven spatial separation of analytes from single droplets entering an ICP off-axis”. Borovinskaya O, Aghaei M, Flamigni L, Hattendorf B, Tanner M, Bogaerts A, Günther D, Journal of analytical atomic spectrometry 29, 262 (2014). http://doi.org/10.1039/c3ja50307k
Abstract: The reproducible temporal separation of ion signals generated from a single multi-element droplet, observed in previous studies, was investigated in detail in this work using an ICPTOFMS with high temporal resolution. It was shown that the signal peak intensities of individual elements temporally shift relative to each other only for droplets moving through the plasma off-axis. The magnitude of these shifts correlated with the vaporization temperatures of the analytes and depended on the radial position of the droplets as well as on the thermal properties and velocity profiles of the carrier gases of the ICP. The occurrence of the signal shifting was explained by a spatial separation of analytes already present in the vapor phase in the ICP from a yet unvaporized residue of the droplet. This separation is most likely driven by anisotropic diffusion of vaporized analytes towards the plasma axis and a radial velocity gradient. The proposed explanation is supported by modeling of the gas velocities inside the ICP and imaging of the atomic and ionic emissions produced from single droplets, whose patterns were sloping towards the center of the torch. The effects observed in these studies are important not only for the fundamental understanding of analyteplasma interactions but have also a direct impact on the signal intensities and stability.
Keywords: A1 Journal article; Plasma Lab for Applications in Sustainability and Medicine – Antwerp (PLASMANT)
Impact Factor: 3.379
Times cited: 18
DOI: 10.1039/c3ja50307k
|
“Occurrence of gas flow rotational motion inside the ICP torch : a computational and experimental study”. Aghaei M, Flamigni L, Lindner H, Günther D, Bogaerts A, Journal of analytical atomic spectrometry 29, 249 (2014). http://doi.org/10.1039/c3ja50302j
Abstract: An inductively coupled plasma, connected to the sampling cone of a mass spectrometer, is computationally investigated. The occurrence of rotational motion of the auxiliary and carrier gas flows is studied. The effects of operating parameters, i.e., applied power and gas flow rates, as well as geometrical parameters, i.e., sampler orifice diameter and injector inlet diameter, are investigated. Our calculations predict that at higher applied power the auxiliary and carrier gas flows inside the torch move more forward to the sampling cone, which is validated experimentally for the auxiliary gas flow, by means of an Elan 6000 ICP-MS. Furthermore, an increase of the gas flow rates can also modify the occurrence of rotational motion. This is especially true for the carrier gas flow rate, which has a more pronounced effect to reduce the backward motion than the flow rates of the auxiliary and cooling gas. Moreover, a larger sampler orifice (e.g., 2 mm instead of 1 mm) reduces the backward flow of the auxiliary gas path lines. Finally, according to our model, an injector inlet of 2 mm diameter causes more rotations in the carrier gas flow than an injector inlet diameter of 1.5 mm, which can be avoided again by changing the operating parameters.
Keywords: A1 Journal article; Plasma Lab for Applications in Sustainability and Medicine – Antwerp (PLASMANT)
Impact Factor: 3.379
Times cited: 21
DOI: 10.1039/c3ja50302j
|
“The effect of the sampling cone position and diameter on the gas flow dynamics in an ICP”. Aghaei M, Lindner H, Bogaerts A, Journal of analytical atomic spectrometry 28, 1485 (2013). http://doi.org/10.1039/c3ja50107h
Abstract: An inductively coupled plasma, connected to a sampling cone of a mass spectrometer, is computationally investigated. The effects of the sampler orifice diameter (ranging from 1 to 2 mm) and distance of the sampler cone from the load coil (ranging from 7 to 17 mm) are studied. An increase in sampler orifice diameter leads to a higher central plasma temperature at the place of the sampler, as well as more efficient gas transfer through the sampler, by reducing the interaction of the plasma gas with the sampling cone. However, the flow velocity at the sampler position is found to be independent of the sampler orifice diameter. Moreover, by changing the sampler orifice diameter, we can control whether only the central gas or also the auxiliary gas can exit through the sampler. Finally, with the increasing distance of the sampler from the load coil, the plasma temperature at the place of the sampler decreases slightly, which might also have consequences for the ion generation and transport through the sampling cone.
Keywords: A1 Journal article; Plasma Lab for Applications in Sustainability and Medicine – Antwerp (PLASMANT)
Impact Factor: 3.379
Times cited: 14
DOI: 10.1039/c3ja50107h
|
“A mobile instrument for in situ scanning macro-XRF investigation of historical paintings”. Alfeld M, Pedroso JV, van Hommes ME, van der Snickt G, Tauber G, Blaas J, Haschke M, Erler K, Dik J, Janssens K, Journal of analytical atomic spectrometry 28, 760 (2013). http://doi.org/10.1039/C3JA30341A
Abstract: Scanning macro-X-ray fluorescence analysis (MA-XRF) is rapidly being established as a technique for the investigation of historical paintings. The elemental distribution images acquired by this method allow for the visualization of hidden paint layers and thus provide insight into the artist's creative process and the painting's conservation history. Due to the lack of a dedicated, commercially available instrument the application of the technique was limited to a few groups that constructed their own instruments. We present the first commercially available XRF scanner for paintings, consisting of an X-ray tube mounted with a Silicon-Drift (SD) detector on a motorized stage to be moved in front of a painting. The scanner is capable of imaging the distribution of the main constituents of surface and sub-surface paint layers in an area of 80 by 60 square centimeters with dwell times below 10 ms and a lateral resolution below 100 mu m. The scanner features for a broad range of elements between Ti (Z = 22) and Mo (Z = 42) a count rate of more than 1000 counts per second (cps)?mass percent and detection limits of 100 ppm for measurements of 1 s duration. Next to a presentation of spectrometric figures of merit, the value of the technique is illustrated through a case study of a painting by Rembrandt's student Govert Flinck (1615-1660).
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
Impact Factor: 3.379
Times cited: 106
DOI: 10.1039/C3JA30341A
|
“Revealing hidden paint layers in oil paintings by means of scanning macro-XRF : a mock-up study based on Rembrandt's “An old man in military costume””. Alfeld M, de Nolf W, Cagno S, Janssens K, et al, Journal of analytical atomic spectrometry 28, 40 (2013). http://doi.org/10.1039/C2JA30119A
Abstract: Over the past several decades the oeuvre of Rembrandt has been the subject of extensive art historical and scientific investigations. One of the most striking features to emerge is his frequent re-use of canvases and panels. The painting An Old Man in Military Costume (78.PB.246), in the collection of the J. Paul Getty Museum, is an example of such a re-used panel. Conventional imaging techniques revealed the presence of a second portrait under the surface portrait, but the details of this hidden portrait have not yet been revealed. Vermilion (HgS) has been identified to have been used nearly exclusively in the flesh tones of the lower painting, suggesting that element-specific XRF imaging might successfully image the hidden portrait. To test this hypothesis, a full-scale mock-up of the painting was created, including a “free impression” of the hidden portrait, reproducing as closely as possible the pigments and paint stratigraphy of the original painting. XRF imaging of the mock-up painting was conducted using three different XRF imaging systems: a mobile X-ray tube based system and two synchrotron-based setups (one equipped with multiple SDDs and one equipped with a Maia detector). The sensitivity, limits of detection and imaging capabilities of each system under the chosen experimental conditions are evaluated and compared. The results indicate that an investigation of the original painting by this method would have an excellent chance of success.
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
Impact Factor: 3.379
DOI: 10.1039/C2JA30119A
|
“Modeling of photon scattering at high X-ray energies : experiment versus simulation”. Vincze L, Vekemans B, Janssens K, Adams F, Journal of analytical atomic spectrometry
T2 –, 15th International Congress on X-Ray Optics and Microanalysis (ICXOM), AUG 24-27, 1998, ANTWERP, BELGIUM 14, 529 (1999). http://doi.org/10.1039/A808040B
Abstract: The use of a detailed Monte Carlo simulation code for X-ray fluorescence spectrometers is demonstrated for calculating the outcome of X-ray scattering experiments in the incident energy range 40-80 keV. The code was validated by comparisons of experimental and simulated spectral distributions in the case of thick, homogeneous samples in which multiple photon scattering occurs with high probability. The experimental spectral distributions were collected at beamline BW5 of HASYLAB, Germany, where a highly energetic, monochromatic synchrotron beam is available. With respect to heterogeneous samples, the code was employed to evaluate the use of Rayleigh and Compton scatter signals for obtaining three dimensional information on the sample dark matrix composition.
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
Impact Factor: 3.379
DOI: 10.1039/A808040B
|
“Analysis of platinum powder by glow discharge mass spectrometry”. van Straaten M, Swenters K, Gijbels R, Verlinden J, Adriaenssens E, Journal of analytical atomic spectrometry 9, 1389 (1994). http://doi.org/10.1039/ja9940901389
Keywords: A1 Journal article; Plasma Lab for Applications in Sustainability and Medicine – Antwerp (PLASMANT)
Impact Factor: 3.466
Times cited: 17
DOI: 10.1039/ja9940901389
|
“Computer simulation of an analytical direct current glow discharge in argon: influence of the cell dimensions on the plasma quantities”. Bogaerts A, Gijbels R, Journal of analytical atomic spectrometry 12, 751 (1997)
Keywords: A1 Journal article; Plasma Lab for Applications in Sustainability and Medicine – Antwerp (PLASMANT)
Impact Factor: 3.379
Times cited: 21
|