“Determination of trace metals in rain water by differential-pulse stripping voltammetry”. Vos L, Komy G, Reggers G, Roekens E, Van Grieken R, Analytica chimica acta 184, 271 (1986). http://doi.org/10.1016/S0003-2670(00)86491-0
Abstract: Differential-pulse stripping voltammetry is applied to measure zinc, cadmium, lead and copper by anodic stripping and selenium(IV) by cathodic stripping in rain water at pH 2; subsequently, at pH 9,1, manganese is measured by anodic stripping on the same portion, and cobalt and nickel are measured in the adsorptive mode after formation of their dimethylglyoximates. The instrumental parameters are optimized. The linear ranges, mutual interferences and detection limits are studied. Excellent accuracy is demonstrated; the standard deviation is around 15% at 2.550 μg l−1 levels. The method is shown to be applicable for rain water.
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
DOI: 10.1016/S0003-2670(00)86491-0
|
“Soil analysis by thin-film energy-dispersive X-ray fluorescence”. Van Grieken R, Van 't dack L, Costa Dantas C, Da Silveira Dantas H, Analytica chimica acta 108, 93 (1979). http://doi.org/10.1016/S0003-2670(01)93044-2
Abstract: Energy-dispersive x-ray fluorescence is advantageous for trace analysis of soils present as thin films. A target thickness of about 2 mg cm-2 provides a compromise between optimal sensitivity and minimal absorption effect or optimal accuracy. Sample preparation involves only suspending the finely ground soil in water and drying this suspension on a thin mylar foil glued on a ring that fits into the x.r.f. spectrometer. The effective sample weight present in the exciting beam area is computed from the scatter peaks, a method that cancels out target heterogeneity problems. High accuracy is demonstrated for many elements in reference soil and rock materials; a precision around 5% and a detection limit around 10 ppm can be achieved. As an illustration, results for 16 trace elements and preliminary interpretation are given for a series of pedologically important soil samples from Brasil.
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
DOI: 10.1016/S0003-2670(01)93044-2
|
“Multi-element analysis of urine by energy-dispersive x-ray fluorescence spectrometry”. Vos L, Robberecht H, Van Dyck P, Van Grieken R, Analytica chimica acta 130, 167 (1981). http://doi.org/10.1016/S0003-2670(01)84161-1
Abstract: For multi-element analysis of human urine, 25-ml samples doped with yttrium as internal standard are evaporated gently and then ashed up to 460°C overnight. The residue is pelletized and analysed by energy-dispersive x-ray fluorescence. Acid addition to facilitate the digestion is not mandatory. Recoveries are nearly quantitative for traces of Fe, Ni, Cu, Zn and Sr, to a lesser extent for lead, but not for arsenic or selenium. The standard deviation per measurement is typically around 6%. The detection limits are such that some 10 elements can be determined simultaneously in normal urine, and possibly more in cases of importance to toxicology or industrial hygiene.
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
DOI: 10.1016/S0003-2670(01)84161-1
|
“Preparation of conducting electrodes from biological samples for multi-element trace analysis by spark-source mass spectrometry or emission spectrometry”. Vos L, Van Grieken R, Analytica chimica acta 164, 83 (1984). http://doi.org/10.1016/S0003-2670(00)85619-6
Abstract: Four decomposition procedures frequently used for biological material (dry ashing, open wet digestion, wet digestion in a teflon bomb and low-temperature ashing) are optimized for the conversion of biological samples to conducting electrodes suitable for multi-element trace determinations by spark-source mass spectrometry or emission spectrometry. The optimized procedures are evaluated with respect to contamination, retention and preconcentration of the trace elements, homogeneity of the electrodes and precision of the final results. Both dry-ashing methods are prone to losses by volatilization; simple dry ashing suffers from contamination problems during electrode preparation. Wet digestion gives better precision; digestion with nitric/sulfuric acids in an open flask is the method of choice for most elements being simpler and giving lower blanks than the bomb method.
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
DOI: 10.1016/S0003-2670(00)85619-6
|
“Identification of inorganic and organic microliths in kidney sections by laser microprobe mass spectrometry”. Verbueken AH, Van Grieken RE, de Broe ME, Wedeen RP, Analytica chimica acta 195, 97 (1987). http://doi.org/10.1016/S0003-2670(00)85653-6
Abstract: Laser microprobe mass spectrometry is used to identify intrarenal microliths; they appear to consist of either oxalate, urate or phosphate. Crystalline and amorphous deposits in rat and human kidney are pin-pointed by the laser beam and their chemical composition determined by mass spectrometry. The method has the potential for wide application in the identification of single organic, inorganic or combination crystals in histological sections.
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation); Laboratory Experimental Medicine and Pediatrics (LEMP)
DOI: 10.1016/S0003-2670(00)85653-6
|
“Internal standard activation analysis of silicon in steel”. van Grieken R, Gijbels R, Speecke A, Hoste J, Analytica chimica acta 43, 381 (1968). http://doi.org/10.1016/S0003-2670(00)89235-1
Abstract: Non-destructive 14-MeV neutron activation analysis for silicon in steel has been applied with 56Mn as internal standard.56Mn is formed from the iron matrix via the 56Fe(n,p)56Mn reaction. Several methods of internal standardisation via56Mn are discussed. The 0.84-MeV photopeak of 56Mn is recommended if steel samples of about the same composition are to be analysed. Chemically analysed steel samples are used as silicon standards. A precision of 0.7% was obtained for an analysis plus standardisation time of 13 min. Special attention was paid to interferences produced by concentration changes of impurity elements. Several possible sources of errors were investigated.
Keywords: A1 Journal article; Plasma Lab for Applications in Sustainability and Medicine – Antwerp (PLASMANT); AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
Impact Factor: 4.513
Times cited: 14
DOI: 10.1016/S0003-2670(00)89235-1
|
“Determination of trace selenium in biological material by preconcentration and X-ray emission spectrometry”. Robberecht H, Van Grieken R, Analytica chimica acta 147, 113 (1983). http://doi.org/10.1016/0003-2670(83)80077-4
Abstract: Selenium is determined in the ng g−1 to μg g−1 range in biological and environmental samples. A wet digestion procedure was optimized with respect to volatility losses and recovery yields, by using 75Se metabolically incorporated into rat organs. Selenium is preconcentrated from the digestion liquid by a two-step reduction with 4 M HCl and ascorbic acid. The colloidal selenium formed is adsorbed on activated carbon and filtered on a Nucleopore membrane for measurement by energy-dispersive x-ray fluorescence. Almost complete recovery was obtained, and the detection limit was 20 ng, corresponding to 10 ng g−1 for a 2-g sample. Biological reference materials were analyzed with satisfactory results, and the accuracy of the method was good.
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
DOI: 10.1016/0003-2670(83)80077-4
|
“Optimization of a simple spotting procedure for x-ray fluorescence analysis of waters”. Smits J, Van Grieken R, Analytica chimica acta 88, 97 (1977). http://doi.org/10.1016/S0003-2670(01)96053-2
Abstract: Several sample preparation methods for waters for energy-dispersive x.r.f. were examined, as well as the influence of sample size on the analytical characteristics. The most satisfactory simple, rapid method proved to be spotting of 1.5 ml of water sample on a Whatman-41 cellulose filter paper provided with a wax ring of 29-mm diameter and evaporating the water with an unheated air stream from underneath. Sensitivities are below 100 p.p.b. for most elements and often below 50 p.p.b. when the optimal secondary fluorescer is used. Accuracy and precision are usually in the 1520 % range. The method is applicable to many dilute aqueous solutions as is illustrated by analysis of industrial water samples and ashed biological material.
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
DOI: 10.1016/S0003-2670(01)96053-2
|
“Sample contamination from a commercial grinding unit”. Van Grieken R, Van de Velde R, Robberecht H, Analytica chimica acta 118, 137 (1980). http://doi.org/10.1016/S0003-2670(01)93724-9
Abstract: The contamination of ground samples by a commercially available Lovibond McCrone Micronizing Mill is discussed. Tracer and weighing experiments showed that abrasion of corundum grinding elements was important, introducing 620 mg of abrasion products per minute of wet grinding. Agate grinding elements were abraded at ⩽6 mg min-1. The abrasion products and grinding elements were analyzed by x-ray fluorescence, spark-source mass spectrometry and neutron activation analysis. Contamination in trace element analysis of geological materials is likely to be negligible for agate grinding elements and, except for a few transition metals, also for corundum grinding elements. Contamination of typical biological samples is significant for a few elements even when agate elements are used, and is absolutely prohibitive for trace analysis when corundum elements are used.
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
DOI: 10.1016/S0003-2670(01)93724-9
|
“Determination of silicon in natural and pollution aerosols by 14-MeV neutron activation analysis”. Gijbels R, Dams R, Analytica chimica acta 63, 369 (1973). http://doi.org/10.1016/S0003-2670(01)82362-X
Abstract: The determination of silicon via the 28Si(n,p)28 Al reaction by means of 14-MeV neutrons is applied to the analysis of pollution and natural aerosols. A Whatman 41 filter (40 cm2) on which airborne particulate material has been collected is compressed into a 3 × 12.7 mm pellet. Standards are prepared in the same way from clean filters spiked with a silicate solution. After a 50-s irradiation and a 75-s decay time, the sample is counted for 2 min with 5 × 5 NaI(Tl) well detector. The 1.779-MeV photopeak of 28Al is measured with a single channel sealer chain or with a multichannel analyser. The reproducibility, sensitivity and liability to interference from other elements were investigated for both counting systems. The homogeneity of the pellets and the filters was checked. The overall precision of one single-channel determination was estimated to be 3.5% after a 24-h high-volume sampling time. Samples collected in urban, industrial and remote areas with concentrations ranging from 0.05 to 15 μg Si m-3 air were analysed and the results are discussed.
Keywords: A1 Journal article; Plasma Lab for Applications in Sustainability and Medicine – Antwerp (PLASMANT)
Impact Factor: 4.513
Times cited: 16
DOI: 10.1016/S0003-2670(01)82362-X
|
“Development of a combi-electrosensor for the detection of phenol by combining photoelectrochemistry and square wave voltammetry”. Neven L, Barich H, Sleegers N, Cánovas R, Debruyne G, De Wael K, Analytica chimica acta 1206, 339732 (2022). http://doi.org/10.1016/J.ACA.2022.339732
Abstract: The high toxicity, endocrine-disrupting effects and low (bio)degradability commonly attributed to phenolic compounds have promoted their recognition as priority toxic pollutants. For this reason, the monitoring of these compounds in industrial, domestic and agricultural streams is crucial to prevent and decrease their toxicity in our daily life. To confront this relevant environmental issue, we propose the use of a combi-electrosensor which combines singlet oxygen (1O2)-based photoelectrochemistry (PEC) with square wave voltammetry (SWV). The high sensitivity of the PEC sensor (being a faster alternative for traditional COD measurements) ensures the detection of nmol L−1 levels of phenolic compounds while the SWV measurements (being faster than the color test kits) allow the differentiation between phenolic compounds. Herein, we report on the development of such a combi-electrosensor for the sensitive and selective detection of phenol (PHOH) in the presence of related phenolic compounds such as hydroquinone (HQ), bisphenol A (BPA), resorcinol (RC) and catechol (CC). The PEC sensor was able to determine the concentration of PHOH in spiked river samples containing only PHOH with a recovery between 96% and 111%. The SWV measurements elucidated the presence of PHOH, HQ and CC in the spiked samples containing multiple phenol compounds. Finally, the practicality of the combi-electrosensor set-up with a dual SPE containing two working electrodes and shared reference and counter electrodes was demonstrated. As a result, the combination of the two techniques is a powerful and valuable tool in the analysis of phenolic samples, since each technique improves the general performance by overcoming the inherent drawbacks that they display independently.
Keywords: A1 Journal article; Antwerp Electrochemical and Analytical Sciences Lab (A-Sense Lab)
DOI: 10.1016/J.ACA.2022.339732
|
“On the precision of oxygen determinations in steel by 14-MeVv neutron activation”. Van Grieken R, Speecke A, Hoste J, Analytica chimica acta 52, 275 (1970). http://doi.org/10.1016/S0003-2670(01)80957-0
Abstract: The precision of the 14-MeV neutron activation determination of oxygen in steel has been examined as a function of the oxygen content for a large number of steel and cast iron samples. The experimental and the statistically expected standard deviations have been compared. In the conditions used, 2.5 counts from 16N were registered for each p.p.m. of oxygen in the samples. The neutron flux was monitored by a simultaneously irradiated oxygen standard.
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
DOI: 10.1016/S0003-2670(01)80957-0
|
“Comparison of preconcentration procedures for trace metals in natural waters”. Smits J, Nelissen J, Van Grieken R, Analytica chimica acta 111, 215 (1979). http://doi.org/10.1016/S0003-2670(01)93264-7
Abstract: The relative merits of eight procedures for preconcentrations of trace metal ions from natural water samples and synthetic solutions are evaluated. Spikes (100 μg l−1 ) of Mn, Co, Zn, Eu, Cs and Ba and the corresponding radioactive tracers were added to batches of drinking water, estuarine water, sea water, ground water, twice-distilled water and ahumic material solution. After equilibration for 25 months, the following techniques were applied: passage through columns of Dowex Al chelating resin and ofsilylated silica gel, filtration through laminate membrane filters and chelating diethylenetriamine cellulose filters, precipitation with sodium diethyldithiocarbamate and l-(2-pyridylazo)-2-naphthol, extraction with ammonium pyrrolidinedithiocarbamate, and chelation by 8-quinolinol (oxine) followed by adsorption on activated carbon. The quantitative characteristics of these techniques and the influence of the water matrix effects are discussed, as well as the applicability for x-ray fluorescence analysis.
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
DOI: 10.1016/S0003-2670(01)93264-7
|
“Determination of precious metals in ores and rocks by thermal neutron activation/\gamma-spectrometry after preconcentration by nickel sulphide fire assay and coprecipitation with tellurium”. Shazali I, Van 't dack L, Gijbels R, Analytica chimica acta 196, 49 (1987). http://doi.org/10.1016/S0003-2670(00)83069-X
Abstract: The six platinum group elements (Ru, Rh, Pd, Os, Ir and Pt) can be determined in geological samples down to the μg kg−1 level, by using nickel sulphide fire assay and neutron activation of the residue ramaining after dissolution of the nickel sulphide button in concentrated hydrochloric acid. Losses for the platinum group elements during this dissolution step are usually insignificant, except when the elements are present at ultra-trace levels. The can be recovered from the filtrate by coprecipitation with tellerium. The latter approach also permits determination of silver, which is significantly lost in the hydrochloric acid treatment (recovery <98% instead of typically ≈ 10%). The coprecipitation with tellurium considerably improves the results for gold (recovery ≈ 95% instead of typically 75%).
Keywords: A1 Journal article; Plasma Lab for Applications in Sustainability and Medicine – Antwerp (PLASMANT)
Impact Factor: 4.513
Times cited: 49
DOI: 10.1016/S0003-2670(00)83069-X
|
“Controlling the mixed potential of polyelectrolyte-coated platinum electrodes for the potentiometric detection of hydrogen peroxide”. Baez JF, Compton M, Chahrati S, Cánovas R, Blondeau P, Andrade FJ, Analytica Chimica Acta 1097, 204 (2020). http://doi.org/10.1016/J.ACA.2019.11.018
Abstract: The use of a Pt electrode coated with a layer of Nafion has been described in previous works as an attractive way to perform the potentiometric detection of hydrogen peroxide. Despite of the attractive features of this approach, the nature of the non-Nernstian response of this system was not properly addressed. In this work, using a mixed potential model, the open circuit potential of the Pt electrode is shown to be under kinetic control of the oxygen reduction reaction (ORR). It is proposed that hydrogen peroxide acts as an oxygenated species that blocks free sites on the Pt surface, interfering with the ORR. Therefore, the effect of the polyelectrolyte coating can be understood in terms of the modulation of the factors that affects the kinetics of the ORR, such as an increase of the H+ concentration, minimization of the effect of the spectator species, etc. Because of the complexity and the lack of models that accurately describe systems with practical applications, this work is not intended to provide a mechanistic but rather a phenomenological view on problem. A general framework to understand the factors that affect the potentiometric response is provided. Experimental evidence showing that the use of polyelectrolyte coatings are a powerful way to control the mixed potential open new ways for the development of robust and simple potentiometric sensors.
Keywords: A1 Journal article; Engineering sciences. Technology; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
Impact Factor: 6.2
DOI: 10.1016/J.ACA.2019.11.018
|
“How perfluoroalkyl substances modify fluorinated self-assembled monolayer architectures : an electrochemical and computational study”. Moro G, Foumthuim CJD, Spinaci M, Martini E, Cimino D, Balliana E, Lieberzeit P, Romano F, Giacometti A, Campos R, De Wael K, Moretto LM, Analytica chimica acta 1204, 339740 (2022). http://doi.org/10.1016/J.ACA.2022.339740
Abstract: There is an urgent need for sensing strategies to screen perfluoroalkyl substances (PFAS) in aqueous matrices. These strategies must be applicable in large-scale monitoring plans to face the ubiquitous use of PFAS, their wide global spread, and their fast evolution towards short-chain, branched molecules. To this aim, the changes in fluorinated self-assembled monolayers (SAM) with different architectures (pinholes/defects-free and with randomized pinholes/defects) were studied upon exposure to both long and short-chain PFAS. The applicability of fluorinated SAM in PFAS sensing was evaluated. Changes in the SAM structures were characterised combining electrochemical impedance spectroscopy and voltam-metric techniques. The experimental data interpretation was supported by molecular dynamics simu-lations to gain a more in-depth understanding of the interaction mechanisms involved. Pinhole/defect-free fluorinated SAM were found to be applicable to long-chain PFAS screening within switch-on sensing strategy, while a switch-off sensing strategy was reported for screening of both short/long-chain PFAS. These strategies confirmed the possibility to play on fluorophilic interactions when designing PFAS screening methods.(c) 2022 Elsevier B.V. All rights reserved.
Keywords: A1 Journal article; Antwerp Electrochemical and Analytical Sciences Lab (A-Sense Lab)
DOI: 10.1016/J.ACA.2022.339740
|
“A study of the weathering of an historic building”. Leysen L, Roekens E, Komy Z, Van Grieken R, Analytica chimica acta 195, 247 (1987). http://doi.org/10.1016/S0003-2670(00)85666-4
Abstract: Various bulk and surface analytical techniques were used to study the chemical deterioration of the 13th-to-15th century limestone cathedral in Mechelen, Belgium. The weathering crust on the walls was found to be rich in sulfate, regardless of the geographic orientation. Nitrate and chloride were only detected in minor amounts in the crust and run-off samples. Attack by gaseous sulfur compounds seems to play a dominant role in the stone deterioration mechanism. Electron microprobe analysis showed predominantly bar-shaped gypsum crystals in the crust, and laser microprobe mass spectrometry showed that carbon seems to be responsible for the blackness of most crust samples. Automated electron microprobe analysis also indicated significant differences in the analytical composition of suspensions in run-off water and in rain-water.
Keywords: A1 Journal article; AXES (Antwerp X-ray Analysis, Electrochemistry and Speciation)
DOI: 10.1016/S0003-2670(00)85666-4
|